Pinhole Detector

ATI offers service and replacement parts for SEMS pinhole detectors.

Applications:
Metal foil
Tin plate

Specifications:
Power Demand                                                                          Approximately 500 VA
Supply Voltage                                                                           230/460 V +/- 10% 60 HZ
                                                                                                     220/440 V +/- 10% 50 HZ
                                                                                                     (Other voltages available upon request)
Temperature Range                                                                  0 to 50 Degrees C
Sensitivity                                                                                   .0005 In. diameter hole
Speed                                                                                          1 to 2000 ft./min.
Strip Thickness:
   Suggested Upper Limit 0.065 In.
   Limited by Shutter 0.185 In.
Strip Width Maximum by Model:*                                          6000 or 60"
Automatic Shutters - Threaded                                              67 inches
                                       Inspect                                                 60 inches
Manual Shutters - Threaded                                                   68 inches
                                  Inspect                                                     62 inches
Minimum which can be inspected                                        12 In. centered
Strip Weave                                                                               Any within width limits.
*Other sizes are available on special request.
Shutter Overlap                                          0.0625 In. minimum at all times. This is normally set to 0.125 In.
Shutter Dead Band                                    +/- .04 In. to +/- .1 In. (factory set approx. .06 In.) Shutter
Shutter Speed                                            240 or 375 inches per min. for 60 HZ Power freq.
                                                                     200 or 312 inches per min for 50 HZ power freq.
                                                                     (factory set for lower speed. Can be changed in the field)
Strip weave speed                                     Must be less than shutter speed
Typical speed-overlap relationships:
Shutter speed                                 240 In. per Min.            to                375 In. per Min.
Normal Overlap                              .125 In.                           to               .185 In.
Minimum Deadband                     +/- .55                             to               +/- .08 In.

OUTPUTS
Relay Contact                               DPDT 2 amp at 115 VAC resistive. Protection required for other loads.                                                                                                             Derate for incandescent lamp loads with high in-rush (1/10 rating).
Closure Delay                                                                Approx. 10 milliseconds after detection
Closure Duration                                                          Adjustable, approx. 10 to 60 MS
Single Pulse                                                                  No holes 0.0V to -0.6V
                                                                                        Pinhole + 8V
Output impedance                                                       Approx. 100 ohms
Minimum load impedance                                         Approx. 10K ohms
Pulse Duration                                                              Approx. 0.8 milliseconds
Pulse Frequency Approx. 500 HZ
Duration of Output                                                       Equal to time Pinhole is in inspection zone.

INTERLOCK (To be supplied by customer)
Function                                                                         Line Start
Type                                                                                Open when line stopped
                                                                                        Closes when line starts
Shutter Retract Switch (To be supplied by customer (Auto Shutters only)
Type                                                                                SPST - Shutters retract to limit when closed
Two (2) zone dual detection